Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Chinese Journal of Radiological Health ; (6): 19-23, 2021.
Article in Chinese | WPRIM | ID: wpr-973712

ABSTRACT

Objective The experiment project was designed to explore the variation of NaI (Tl) gamma spectrometer channels with environmental temperature. 60Co and 152Eu were used to verify the reliability of the correction methods. Methods Two correction methods were applicated, which were curve fitting correction method and known measurement peak correction method. Results The experimental results showed that temperature changes had an effect on NaI (Tl) measured spectra peak. The relative peak will shift to the right at 5℃ by 9.6%, and to the left at 60℃ by 16%, with the reference temperature set at 25℃. The two methods are based on the channel change due to temperature changes, and they could effectively correct the temperature peak-drift. Conclusion In order to make the measured spectrum information accurate and reliable in field monitoring, it is suggested to monitor the environmental temperature so as to correct the measured data.

SELECTION OF CITATIONS
SEARCH DETAIL